The calibration source is an acetylene gas cell C2H2 with an LED behind it; the OSA will display the spectral absorption rate of the acetylene gas.
This absorption characteristic is used to re-center the wavelength calibration table to a known peak during the calibration procedure.
The sideband width is not specified; only the peak wavelength is specified.