What is the WT5000 integration time accuracy? If integration is manually stopped in between a whole second count, does the WT5000 integrate until the second is complete?

The integration time has the accuracy of the real-time clock, not the clock of the AD converter. This means that the uncertainty is +/- 0.02%.

When integration is manually stopped between a whole second count, the displayed data will be the last updated data. It will be held with the data up to 1 second ago.