Is it Possible to Synchronize More Than 3 DL750?

Yes, it is possible to synchronize more than 3 DL750's by using the trigger signal, in a one-to-one connection method, as long as the limits of the delay time is not exceeded. 
 
Example Configuration:
DL750 [TRIG OUT]――>[TRIG IN] DL750 [TRIG OUT]――> [TRIG IN] DL750 [TRIG OUT]――>[TRIG IN] DL750
 
 
For time sensitive applications, it is possible to connect the instruments as shown below. However, please shorten the wiring length between the DL750's as short as possible (for example, using T-adaptor).
  
DL750[TRIG OUT]―┐
                               ├> [TRIG IN] DL750
                               ├> [TRIG IN] DL750
                               ├> [TRIG IN] DL750
                               └> [TRIG IN] DL750
 
The attached document explains the delay time in synchronized operation, which also applies to the DL750P and SL1400.

Related Products & Solutions

Data Acquisition (DAQ)

Data acquisition (DAQ) is the measurement, recording, analyzing, and presentation of real world phenomena. It includes electrical measurements such as voltage, current, power as well as measurements through sensors and transducers including temperature, pressure, strain, vibration, and more.

High Speed Data Acquisition

Yokogawa high speed data acquisition systems deliver industry leading isolation, bit resolution, sampling rate, and memory depth, with independent channel hardware and easy to use software.

Oscilloscopes

Capture, analyze, and troubleshoot electrical or real-world physical signals. Oscilloscopes observe the change of electrical signals over time, continuously graphed on a display as voltage or amplitude vs. time.

ScopeCorders

A ScopeCorder is an instrument combining a mixed signal oscilloscope and portable data acquisition recorder into a modular platform designed to capture both high-speed transients and low-speed trends. Yokogawa’s ScopeCorder product family provides flexible and high-performance multi-channel test instruments by combining a variety of signal conditioning input modules, onboard calculations, and deep data acquisition storage into an off-the-shelf data acquisition solution.

 
Top