PP COMPRESSION setting lost on DL850 ScopeCorder?

When I load a .SET file, I noticed that my PP COMPRESSION setting for WDF File Compression is disabled, why?

This is an intentional design. The settings for the saved data history (One / All) are not saved in the setting file, and the settings for P-P compression (ON / OFF) are not saved in the setting file.

How Do I Compress PP on DL850 ScopeCorder?

You could leave it OFF and save large files and do the PP Compression as post-process within Xviewer. 

Another suggestion to do the PP compression:

 1. Turn ON ENVELOPE Mode

2. Then set chassis sample-rate to the one low-rate that you want 

3. (these settings will save in the SET file)

ENVELOPE mode runs the module at full-speed, and uses PP COMPRESSION to discard the points to achieve the new sample-rate (low) that you want.

Related Products & Solutions

DL850/DL850V ScopeCorder

Combining the capabilities of an Oscilloscope and the characteristics of a Data Acquisition Recorder, the Yokogawa Scopecorder is a versatile tool for engineers making measurements in the industrial, transportation and energy sectors.

DL850E/DL850EV ScopeCorder

An all-in-one electromechanical measurement system offering the deepest measurement and analysis features to turn data into insights.

Precision Making

Top