Sample code for pulling just motor readings through command prompts on WT1800E

Here I have set up a sample code to make a list with 5 values in it.

I chose the values from the motor evaluation function list. 

 

:numeric:normal:number 5        

Here I made the number of values in the list 5

:NUMERIC:NORMAL:ITEM1 SPEed

 Here I made the first item speed 

:NUMERIC:NORMAL:ITEM2 Torq

 Here I made the second item torq

:NUMERIC:NORMAL:ITEM3 sync

 “”

:NUMERIC:NORMAL:ITEM4 Slip 

“”

:NUMERIC:NORMAL:ITEM5 PM

 “”

:numeric:normal? 

Here I returned the item setting I just set to be sure they were right 

:numeric:normal:value?

 Here I queried the values of those items in my list

 

Here are the corresponding responses:

 

<--- :NUMERIC:NORMAL:ITEM3 Torq ( 

<--- :NUMERIC:NORMAL:ITEM1 SPEed ( 1 ms )

<--- :NUMERIC:NORMAL:ITEM2 Torq ( 1 ms )

<--- :NUMERIC:NORMAL:ITEM3 sync ( 1 ms )

<--- :NUMERIC:NORMAL:ITEM4 SLip ( 30 ms )

<--- :NUMERIC:NORMAL:ITEM5 PM ( 1 ms )

<--- :numeric:normal? ( 1 ms )

---> :NUM:NUM 5;ITEM1 SPE;ITEM2 TORQ;ITEM3 SYNC;ITEM4 SLIP;ITEM5 PM ( 2 ms )

<--- :numeric:normal:value? ( 1 ms )

---> -0.001E+00,-0.002E+00,2.5496E+03,100.000E+00,0.000E+00 ( 2 ms )

Related Products & Solutions

WT1800E - High-Performance

  • Power, Harmonics (THD) Measurements (up to 500th order)
  • Electrical & Mechanical Efficiency of Inverters and Motors in EV and HEV

 

 

WT1800R High-Performance Power Analyzer

  • Power, harmonics (THD) measurements (up to 500th order)
  • Electrical and mechanical efficiency of inverters and motors in EV and HEV

Power Analyzers and Power Meters

Measure characteristics of devices that generate, transform or consume electricity. Also called power meters or wattmeters, these devices measure parameters such as true power (watts), power factor, harmonics, and efficiency.

Precision Making

Top