ACTION ON TRIGGER disables itself after POWER CYCLE on ScopeCorder - why?

Why does the “ACTION ON TRIGGER” configuration not stay ON when power is cycled, and it has to be configured again or the setup loaded?

It does clear with power-cycle. To fix this, press: UTILITY, PREFERENCE, and then POWER ON ACTION. Set it for ACTION to ON and set it to START if you wish for it to also recover from a power-outage and it will press the START button for itself, so to speak.

Related Products & Solutions

DL850/DL850V ScopeCorder

Combining the capabilities of an Oscilloscope and the characteristics of a Data Acquisition Recorder, the Yokogawa Scopecorder is a versatile tool for engineers making measurements in the industrial, transportation and energy sectors.

DL850E/DL850EV ScopeCorder

An all-in-one electromechanical measurement system offering the deepest measurement and analysis features to turn data into insights.

ScopeCorders

  • Flexible, high-performance Yokogawa Test&Measurement ScopeCorder
  • Modular platform combines mixed signal oscilloscope and portable data acquisition recorder 
  • Captures high-speed transients and low-speed trends 

Precision Making

Top