Is it avalable that series or parallel connection of source/measure unit GS610/GS820?

The GS610/GS820 output is insulated from the cover case. Therefore, if you want to amplify the output voltage more than 1 unit (or 1 ch) you must connect the system in series. If you want to increase the output current by more than 1 unit (or 1 ch) you must connect the system in parallel. (Provided that it is within a range of maximum voltage/current or those limiter)
 
However, in case of a parallel connection in voltage output or series connection in current, a protection circuit may prevent normal operation and accuracy. Yokogawa cannot guarantee the operation or accuracy when there is a parallel connection in the voltage terminals or a series connection in the current terminals.
 
The attached file provides a detailed explanation for each case.

Related Products & Solutions

GS200 DC Voltage / Current Source

The GS200 is a low voltage/current DC source with high accuracy, high stability, and high resolution.

GS610 Source Measure Unit

The GS610 is made up of a programmable current and voltage source, a voltmeter, and an ammeter. Each function can be combined into numerous operation modes.

GS820 Multi Channel Source Measure Unit

The GS820 is a highly accurate and highly functional 2-channel programmable DC voltage/current source that incorporates voltage/current generation and measurement function.

Signal Generators, Sources and Supplies

Power Meters and data acquisition systems are used in many applications in the test and measurement industry. However, other situations require standalone devices that can source a variety of precise voltage and current waveforms, as well as allow integration into a larger test system. 

Source Measure Units

A Source Measurement Unit (SMU) is a flexible test instrument which is capable of both sourcing and measuring voltage and/or current at the same time.

 
Top