Can I Convert AQ6370 Waveform Data from a Binary to a ASCII .CSV File?

It is possible to convert an AQ6370 Binary file to ASCII .CSV file using the AQ6370 Viewer Software. Please keep in mind there are different AQ637X series OSA Viewer software will load specific binary file extensions.  ...

item: 1129  |  http://tmi.yokogawa.com/ru/service-support/faqs/1129/

On the AQ637X Series OSA Why is the Actual Resolution Not the Set Value?

The resolution is controlled by 3 factors: Input Slit Detraction Grating  Output Slit The resolution setting controls the output slit disk. This disk has a fixed number of slits in it. Since the setting only control on...

item: 1057  |  http://tmi.yokogawa.com/ru/service-support/faqs/1057/

What is the Largest Core Diameter Fiber I can use in the AQ6373, AQ6375 and AQ6370 series OSAs?

For the AQ6373 you can use a maximum 800 micron core fiber that must be terminated in a standard FC connector.   For the AQ6375 and AQ6370 series OSA you can use a maximum 200 micron core fiber which, again must be term...

item: 1112  |  http://tmi.yokogawa.com/ru/service-support/faqs/1112/

Is the RMS Method for Spectrum Width Calculation on the AQ6370C FOTP-127 Compliant?

The AQ6370C OSA was developed in accordance with the IEC standard: WG4 (IEC 61280-1-3) Test procedures for general communication subsystems -Central wavelength and spectral width measurement. We cannot guarantee the ...

item: 1308  |  http://tmi.yokogawa.com/ru/service-support/faqs/1308/

How do I Resolve the Warning Message:142 WL Calibration Failed?

If you are performing wavelength calibration using the instrument's built-in light source, the warning message is an indication that the monochromator needs to be repaired. Please visit our Contact site to locate and c...

item: 401  |  http://tmi.yokogawa.com/ru/service-support/faqs/401/

Is the AQ6370 Viewer Software Compatible with Windows 7?

Yes, the AQ6370 Viewer Software is compatible with Windows 7 32-bit and 64-bit. Note: Please make sure the correct USB driver is installed for either Windows 7 32-bit or 64-bit....

item: 991  |  http://tmi.yokogawa.com/ru/service-support/faqs/991/

Why Does my AQ637X OSA Freeze Every so Often?

The AQ637X OSA has a feature called Auto Offset that it performs every 10 minutes. What this feature does is zero out the OP amps in the OSA to ensure constant level accuracy. This feature will however, stop the OSA from...

item: 1081  |  http://tmi.yokogawa.com/ru/service-support/faqs/1081/

Why Can't the AQ637X Series OSA Detect or Display an External Display?

If the AQ637X series OSA is already running and you attempt to connect an external display, the OSA will not automatically detect the external display. The OSA will need to be re-booted before it can recognize and displa...

item: 174  |  http://tmi.yokogawa.com/ru/service-support/faqs/174/

Does Yokogawa Sell Optical Fibers?

No, unfortunately due to the number of current fiber combinations, we do not stock optical fibers. There are many companies that specialize in optical fiber cords....

item: 1078  |  http://tmi.yokogawa.com/ru/service-support/faqs/1078/