What is the Limit for Each Module That can be Placed in a DL850?
The number of each module that can be placed in a DL850 is limited to a maximum of: Voltage Scanner Module: 8 CAN Bus Monitoring Module: 2 (must be installed in slot 7 & 8) Logic Input Module: 8 100 MS/sec Module: 4 (mu...
item: 709 | http://tmi.yokogawa.com/tw/service-support/faqs/709/
Why Can't the SL1000 Acquisition Software Detect my USB Connection?
Even though the correct USB drivers have been installed, and the device displays correctly in the Windows Device Manager Window, the SL1000 Acquisition Software fails to detect the SL1000 main unit. Solutions: Verify th...
item: 675 | http://tmi.yokogawa.com/tw/service-support/faqs/675/
Why Does the Trigger Activate Repeatedly When the Detection Level is set to "H Active" and External Trigger Input is "H" for the WE7251/WE7271 Digitizer Module?
The external trigger input (TRIG IN) as well as external input/output (EXT. I/O) trigger signal becomes the level signal. The WE7251/WE7271 digitizer module will detect the bus trigger by that level. As long as the exte...
item: 1174 | http://tmi.yokogawa.com/tw/service-support/faqs/1174/
How Long is the WE7251/WE7271/WE7272/WE7275 Digitizer Module's Record Length?
For high-speed data recording in trigger mode/gate mode, the digitizer module is equipped with internal memory to store measurement data in order to record reliable data without receiving any effects of the load conditio...
item: 1148 | http://tmi.yokogawa.com/tw/service-support/faqs/1148/
How Much Lag Time is There When Inputting or Outting a Signal from the EXT Terminal of the WE7000 Measurement Station?
The lag time from the external terminal to the bus trigger / time base is as follows:When inputting from TRIG IN:TRIG IN -> bus trigger signal : 40 ns (typical value)TRIG IN -> time base signal : 25 ns (typical value)Whe...
item: 1139 | http://tmi.yokogawa.com/tw/service-support/faqs/1139/
Is the Data Assignment the Same When the Logic Input Values are Saved in ASCII Format on the DL708/DL708E and the DL2700?
DL708/DL708E: The ON/OFF conditions of bit signals from A1 to A8 and B1 to B8 are expressed as a 16-bit value. The 16-bit value is then converted to a decimal value and the result is saved. The value is written as a numb...
item: 1220 | http://tmi.yokogawa.com/tw/service-support/faqs/1220/
Can the Width of the Harmonic Analysis Window be Increased Beyond 16?
In order to analyze the harmonics by the PLL synchronization, the maximum possible window width is automatically determined depending on the frequency of the fundamental wave of the input signal (fundamental frequency f)...
item: 1135 | http://tmi.yokogawa.com/tw/service-support/faqs/1135/
What does the Cursor Value on the AQ7275 Measure?
The cursor value displays the level (dB) of the signal that is above the noise floor of the connected optical port. The noise floor is the dynamic range. This lets you know if you are approaching the maximum distance tha...
item: 1054 | http://tmi.yokogawa.com/tw/service-support/faqs/1054/
What are the Minimum Configuration Requirements for the WT3000 to Work with the Flicker Software?
To use the WT3000 with the Flicker Software, the WT3000 must have the following options: /G6 option - Advanced Calculations /FL option - Flicker Measurements 1 to 3 elements - 30A Input Module Note: The 2A input mod...
item: 1051 | http://tmi.yokogawa.com/tw/service-support/faqs/1051/
Why are the Measured Values for VAR and ΣVAR Unstable?
Check the Synch Source and Frequency Filter settings When a single-phase signal being measured fluctuates around power factor of 1.Slight fluctuations in the measured values of voltage, current, and power can cause a la...
item: 1268 | http://tmi.yokogawa.com/tw/service-support/faqs/1268/