What is the Correct AC Coupling Test for the 720210 Analog Voltage Input Module?

The correct AC coupling test for the 720210 Analog Voltage input module must be performed at 20 Hz.
 
The service manual, SMDL850-01EN_010 1st Ed, suggests that you perform the AC COUPLING test at 15 Hz. This is a typo and incorrect, the test must be performed at 20 Hz.

Additional Information
  • Confirmation: The QIS720210-E_010 has the correct procedure outline for the AC COUPLING TEST at 20Hz.
  • Confirmation: Yokogawa Corp of America Cal Lab using Intercal METCAL script performs this test at 20 Hz.

Related Products & Solutions

SL1000 High Channel Count ScopeCorder

A headless ScopeCorder system capable of connecting multiple chassis into one PC-based system.

 

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

  • Yokogawa high-speed data acquisition systems
  • Industry-leading isolation, bit resolution, sampling rate, and memory depth
  • Independent channel hardware and easy to use software

Oscilloscopes

  • Digital oscilloscopes
  • High-speed sampling
  • Range of bandwidths for electronic device design and development
  • Advantages of oscilloscope and multi-channel data recorder

ScopeCorders

  • Flexible, high-performance Yokogawa ScopeCorder
  • Modular platform combines mixed signal oscilloscope and portable data acquisition recorder 
  • Captures high-speed transients and low-speed trends 

 

Precision Making

Top