High Speed Stray Light Reduction Function
New High Speed Stray Light Reduction Function!
(SWITCH Sweep Function)
Measurement with stray light reduction is faster!
A high speed stray light* reduction function (SWITCH sweep function) has been added, speeding up measurement times. When compared to the CHOP function used in conventional models, the stray light reduction function is Approximately 2.4 times faster.
Measurement conditions: SPAN: 0.5 nm, RES: 10 pm, SENS: HIGH1/SW, SAMPLE: AUTO
Approximately 2.4 times faster measuring speed compared to the CHOP Sweep function.
* Stray light originates from reflection and scattering on the surfaces of optical components inside optical instruments and their holders. As it is from factors other than normal refraction and reflection, it can be harmful to image formation.
With high performance capabilities and a wavelength resolution of 10 picometer or less, the AQ6319 is designed for advanced research and development applications. With a wavelength range or 600 to 1700 nm this OSA offers an excellent combination of wavelength range and resolution.