Can the WT1800/WT1800E measure Total Demand Distortion (TDD)?

Yes, the WT1800 can measure TDD using the following user-defined equation:

ITHD(E1)*IK(E1,OR1)/IL

IL=maximum demand load current. IL should be a constant entered by the customer based on IEEE 519 standards. The above equation is written for element 1.

Related Products & Solutions

WT1800 High Performance Power Analyzer

The WT1800 Power Analyzer offers maximum flexibility with up to six wattmeter elements, high bandwidths, simultaneous high speed digitizing, and wide voltage and current ranges. Although this model is still available, the newer WT1800E offers higher accuracy.

WT1800E - High-Performance

  • Up to 6 wattmeters
  • 0.05% of rdg + 0.05% of rng
  • DC, 0.1Hz to 1MHz
  • 2 motor inputs
  • Harmonics to 500th order

WT1800R High-Performance Power Analyzer

  • Up to 6 wattmeters
  • 0.05% of rdg + 0.05% of rng
  • DC, 0.1Hz to 1MHz
  • 2 motor inputs
  • Harmonics to 500th order