Sample code for pulling just motor readings through command prompts on WT1800E

Here I have set up a sample code to make a list with 5 values in it.

I chose the values from the motor evaluation function list. 

 

:numeric:normal:number 5        

Here I made the number of values in the list 5

:NUMERIC:NORMAL:ITEM1 SPEed

 Here I made the first item speed 

:NUMERIC:NORMAL:ITEM2 Torq

 Here I made the second item torq

:NUMERIC:NORMAL:ITEM3 sync

 “”

:NUMERIC:NORMAL:ITEM4 Slip 

“”

:NUMERIC:NORMAL:ITEM5 PM

 “”

:numeric:normal? 

Here I returned the item setting I just set to be sure they were right 

:numeric:normal:value?

 Here I queried the values of those items in my list

 

Here are the corresponding responses:

 

<--- :NUMERIC:NORMAL:ITEM3 Torq ( 

<--- :NUMERIC:NORMAL:ITEM1 SPEed ( 1 ms )

<--- :NUMERIC:NORMAL:ITEM2 Torq ( 1 ms )

<--- :NUMERIC:NORMAL:ITEM3 sync ( 1 ms )

<--- :NUMERIC:NORMAL:ITEM4 SLip ( 30 ms )

<--- :NUMERIC:NORMAL:ITEM5 PM ( 1 ms )

<--- :numeric:normal? ( 1 ms )

---> :NUM:NUM 5;ITEM1 SPE;ITEM2 TORQ;ITEM3 SYNC;ITEM4 SLIP;ITEM5 PM ( 2 ms )

<--- :numeric:normal:value? ( 1 ms )

---> -0.001E+00,-0.002E+00,2.5496E+03,100.000E+00,0.000E+00 ( 2 ms )

Related Products & Solutions

Power Analyzers and Power Meters

Ensure compliance and optimize efficiency with Yokogawa power analyzers. Achieve unmatched accuracy in watts, power factor, harmonics, and efficiency—design greener systems and validate performance confidently.

WT1800E - High-Performance

  • Up to 6 wattmeters
  • 0.05% of rdg + 0.05% of rng
  • DC, 0.1Hz to 1MHz
  • 2 motor inputs
  • Harmonics to 500th order

WT1800R High-Performance

  • Up to 6 wattmeters
  • 0.05% of rdg + 0.05% of rng
  • DC, 0.1Hz to 1MHz
  • 2 motor inputs
  • Harmonics to 500th order
  •  

Precision Making

Top