Why are the Thermocouple Codes Shown on the Type 2553 DC Voltage/Current Standard Different From the Service Manual?

The specifications for the Type 2553 DC Voltage/Current Standard was changed on June 1982. The service manual was not rewritten to reflect these changes. There are two types of 2553:          1979/4 -- 1982/5         255...

item: 904  |  http://tmi.yokogawa.com/au/service-support/faqs/904/

What is Hi-Resolution Mode?

High Resolution mode can be used to remove high frequency noise and increase vertical resolution. It achieves this by increasing the number of effective bits per data to 12-bits through digital and bandwidth filters. Th...

item: 1076  |  http://tmi.yokogawa.com/au/service-support/faqs/1076/

Power Analyzer Accuracy and Basic Uncertainty Calculator

The Power Analyzer Accuracy and Basic Uncertainty Calculator can be used to determine the uncertainty in voltage, current, and active power (watts) measurement values for various frequency ranges and wiring systems. To u...

item: 417  |  http://tmi.yokogawa.com/au/service-support/faqs/417/

Can the External Clock and External Trigger Functions on the DL1500 Series be Specified or Used Simultaneously?

Yes, there is no restriction on specifying or using the two functions simultaneously. ...

item: 1201  |  http://tmi.yokogawa.com/au/service-support/faqs/1201/

What is Decimation and How do I use Decimation in Xviewer?

Decimation is a technique used to reduce the total number of samples. You can use Xviewer to perform decimation on your waveform data files that have the WDF/WVF/ASCII CSV extension format. Decimation reduces the total n...

item: 1029  |  http://tmi.yokogawa.com/au/service-support/faqs/1029/

Why Doesn't Microsoft Snapshot Viewer Open Yokogawa .SNP Files?

The Yokogawa .SNP file is a proprietary Yokogawa file format for saving waveform snapshots. This file has no relation to the Microsoft Snapshot Viewer program. The .SNP file can only be viewed on the oscilloscope. ...

item: 37  |  http://tmi.yokogawa.com/au/service-support/faqs/37/

How Can I Effectively Achieve a Sample Rate Less Than 5 Sa/sec on the DL850?

The lowest possible device setting for sample rate on the DL850 is 5 Sa/sec. There are however, different methods that will allow you to achieve much lower sample rates which includes the use of an external clock, decim...

item: 1003  |  http://tmi.yokogawa.com/au/service-support/faqs/1003/

Can a USB to Serial Adapter/Converter be Used with a RS232C (DB25) Port on the WT210/WT230?

Although Yokogawa does not have any official recommendation for USB to Serial converters, it is possible to use some USB to RS232 DB9 Adapter Cables. If you choose to use a laptop PC without RS232C port, please use a PC...

item: 1079  |  http://tmi.yokogawa.com/au/service-support/faqs/1079/

What is the Chopped Light Mode?

The Chopped Light mode is used: To increase measurement level sensitivity by cutting the high frequency noise To detect only the light that the LS emitted, effective for free space measurement. In free space measurement...

item: 297  |  http://tmi.yokogawa.com/au/service-support/faqs/297/

Why is the Theoretical Calculated Power Factor and Phase Angle Different Than the Measured Values?

When measuring input signals of distorted waves, signals that are DC-offset or signals that include superimposed harmonic components, will result in different values for power factor and phase angle than those expected w...

item: 1263  |  http://tmi.yokogawa.com/au/service-support/faqs/1263/