If you send "MEASURE:VALUE?" and read out the measured data, 3 data are output as shown in the following example:
Example Output: 4.083E+00,50.002E+00,202.65E+00
"Output Format for Normal Measurement Mode," shows that the order of data output is: V1, A1, W1 etc. Therefore in the example, the first data is V1=4.083E+00, the second data is A1=50.002E+00, and the last data is W1=202.65E+00.
In the "MEASURE:ITEM:NORMAL:PRESET XXX" communications command which was sent first, ALL, DEFAULT1, and DEFAULT2 are available in addition to CLEAR, so you can output only the specified items of the measured values. Also, for example, if you want to output the VA item in addition to "DEFAULT1," you can send "MEASURE:ITEM:NORMAL:PRESET DEFAULT1", then send "MEASURE:ITEM:NORMAL:VA:ELEMENT1 ON" to add it to the DEFAULT1 item. On the other hand, if you want to remove W on element 1 from the "DEFAULT1" setting item, you can send "MEASURE:ITEM:NORMAL:PRESET DEFAULT1", then send "MEASURE:ITEM:NORMAL:W:ELEMENT1 OFF." In this way you can specify individual items for output by adding output items to or taking output items away from the CLEAR, ALL, DEFAULT1, and DEFAULT2 preset settings.
For more information, please refer to the manuals:
Die digitalen Leistungsmessgeräte der Serie WT2000 wurden im Hinblick auf ein optimales Preis/Leistungsverhältnis (Bandbreite, Genauigkeit, Messgeschwindigkeit und Störunterdrückung) entwickelt. Durch den großen Funktionsumfang lassen sich diese Messgeräte in unterschiedlichen Anwendungsbereichen einsetzen.
Für einen effizienten Energie-Einsatz wird eine genauere und zuverlässigere Leistungsmessung immer wichtiger. Einschwingvorgänge, STANDBY-Modus, Transformatoren, Tests und verzerrte Signale durch Inverter, Motoren, Beleuchtungsschaltungen, Stromversorgungen etc., erfordern stabile, vertrauenswürdige und normgerechte Messungen.