How do I Update the Voltage and Current Range on the Instrument Using Pre-IEEE488.2 Command Types?

To change the voltage range on element 1 to the 30 V range, send the "RV1,4" command to the measuring instrument.
To change the current range on element 1 to the 1 A range, send the "RA1,5" command to the measuring instrument.
To set the ranges on all elements at once, use: Voltage global setting "RV0,4" or Current global setting "RA0,5"

After entering settings, confirm the settings by sending the following commands:

  • To query the voltage on element 1: "RV1?"
  • To query the current on element 1: "RA1?" 

For more information on the correspondence between voltage values, current values, and parameters, please refer to:

  • WT1010 Digital Power Meter User's Manual
    Page App1-11
  • WT1030/WT1030M Digital Power Meter User's Manual
    Page App1-13
  • WT2010 Digital Power Meter User's Manual
    Pages App1-14 and App1-15
  • WT2030 Digital Power Meter User's Manual
    Page App1-15

Related Products & Solutions

WT2010/WT2030 Digitale Leistungsmessgeräte

Die digitalen Leistungsmessgeräte der Serie WT2000 wurden im Hinblick auf ein optimales Preis/Leistungsverhältnis (Bandbreite, Genauigkeit, Messgeschwindigkeit und Störunterdrückung) entwickelt. Durch den großen Funktionsumfang lassen sich diese Messgeräte in unterschiedlichen Anwendungsbereichen einsetzen.

Digitale Leistungsanalysatoren

Für einen effizienten Energie-Einsatz wird eine genauere und zuverlässigere Leistungsmessung immer wichtiger. Einschwingvorgänge, STANDBY-Modus, Transformatoren, Tests und verzerrte Signale durch Inverter, Motoren, Beleuchtungsschaltungen, Stromversorgungen etc., erfordern stabile, vertrauenswürdige und normgerechte Messungen.

Precision Making