702928 PBD0200 Differential Probe 20 V / 200 MHz

The 702928 Differential Probe delivers accurate differential voltage measurement up to ±20 V with 200 MHz bandwidth, automatic probe interface support, and low noise performance enabling easy automotive bus signal captures.

When connected to DLM3000/3000HD and DLM5000/5000HD series Yokogawa oscilloscopes, the probe is automatically powered, fully recognized, and is configured to prevent setup errors. 

The 702928 is ideal for capturing CAN and CAN FD waveforms.

Frequency Band (-3 dB) DC to 200 MHz
Attenuation Ratio 10:1
DC gain accuracy ±1%
Differential input voltage range
(between +/- terminals)
±20 V (DC + ACpeak)
Maximum input voltage (to ground) ±60 V DC, ±42 V ACpeak
CMRR (typical) -80 dB (60 Hz), -70 dB (100 kHz),
-50 dB (10 MHz), -20 dB (100 MHz)
Input equivalent noise (typical) 4 mVrms or less
Input capacitance (to ground) (typical) 2 pF
Input resistance (to ground) (typical) 500 kΩ
Output Connector Type Yokogawa dedicated probe interface
Operating temperature range +5°C to +40°C
Operating humidity range 20% to 80% RH (no condensation)
Probe length Approx. 1.2 m
Weight Approx. 90 g