How do I Monitor the Start/Stop Status of the DL716 Digital Oscilloscope Using Communication Commands?

It is possible to monitor the START/STOP status of the DL716 Digital Oscilloscope using communication commands. To do this, you will have to send the ":STATus:CONDition?" (without quotations) command to the instrument, in order to query the status of the Condition Register. The instrument will return a decimal number and Bit-0 will display either 1(Start) or 0 (Stop). Since the return value is a decimal number, you cannot perform bit masking.

As an example, if the DL716 is in START and measurement parameter function is ON, Bit-8 and Bit-0 will display "1." The decimal value for Bit-8 is 128 and for Bit-1 is 1, so the return value of the ":STATus:CONDition?" command is 129.

You cannot use the Extended Event Register to monitor the START/STOP status. The reason for this is because the Extended Event Register is a latch register, and you cannot monitor bit changes at the end of data acquisition from a latch register.

Related Products & Solutions

Anwendungssoftware für Oszilloskope

Mit der Oszilloskop-Applikationssoftware von Yokogawa können Sie Ihre Daten betrachten, exportieren, analysieren und managen.

Oszilloskope

Erstklassige Digital-Oszilloskope von Yokogawa: Die Digital-Oszilloskope zeichnen sich durch eine hohe Abtastrate und eine große Auswahl von Bandbreiten aus und eignen sich für einen Einsatz im Design und der Entwicklung von elektronischen Bauteilen und Geräten. Die ScopeCorder kombinieren die Vorteile sowohl von einem Digital-Oszilloskop, als auch von einem mehrkanaligen Datenrecorder.