Why Won't the Measured Values of the Fundamental Wave Components and the Bar Graph of the Harmonics Display on the WT3000?

When the selected data update cycle of the fundamental wave is shorter than the width of the analysis window determined by the fundamental frequency (cycle of the fundamental wave), measurement is not performed and no data is displayed. Change the currently selected data update rate to something longer.

For harmonic measurement of a 50 Hz distorted wave signal, the fundamental frequency is 50 Hz and the width of the analysis window is ten waves, so the data measurement interval is 200 ms. Since (data measurement interval + data computation interval) = approximately 300 ms or more, select a data update rate of 500 ms or more.

 

Related Products & Solutions

WT3000 Präzisions-Leistungsanalysator

Das Flaggschiff unter den Leistungsmessinstrumenten und Referenzgerät bei anspruchsvollen Wirkungsgradmessungen. Exzellente Linearität und Langzeitstabilität zeichnen das WT3000 aus. Ideal auch für Trafo-Verlustmessungen bei extrem kleinen Leistungsfaktoren. Besonderheiten sind weiterhin Transienten-Aufzeichnung und FFT Analyse. 

 

Digitale Leistungsanalysatoren

Für einen effizienten Energie-Einsatz wird eine genauere und zuverlässigere Leistungsmessung immer wichtiger. Einschwingvorgänge, STANDBY-Modus, Transformatoren, Tests und verzerrte Signale durch Inverter, Motoren, Beleuchtungsschaltungen, Stromversorgungen etc., erfordern stabile, vertrauenswürdige und normgerechte Messungen.