Error(Code:539) Occurs When Loading *.WDF Files to the DL850 ScopeCorder

Error(Code:539)Press ESC Key to Close.

Module configuration is not matched,so it couldn’t loaded.
Configuration of saved data can see by File property.

 

The following error is encountered when the end-user attempts to re-load a *.WDF file to the DL850 ScopeCorder display/acquisition memory from media and the module is installed in the incorrect slot.
This error can be corrected by the following procedures:

  1. Press the FILE key
  2. Press the Utility soft button
  3. Highlight the file and select File Property
  4. Install the modules in the correct slots according to the File Property information
Please download the attached images to view the screenshots of the error code and file property list.