Items | TA320 Advantage | |
Rotational Jitter of Mirror | Time Variation Display | → Analysis of Rotational Deviation |
Accuracy of Mirror Surface |
Continuous measurement |
→ Possibility to evaluate the entire surface of the mirror |
Fixed Accuracy of mirror |
HIgh Resolution |
→ Evaluation of high resoulution |
The reflection at the point A on a mirror surface is detected by Detector A, and the other reflection at point B on the same surface is detected by Detector B.
Rotational jitter of the motor, which is fixed with polygon mirror, is evaluated with a Time-interval measurement (A to B) using the TA320. Transient analysis of motor is available using the time variation display of the TA320.
TA320 Setting
Example
Long term jitter indicates rotational jitter of the motor.
Short term jitter measures the data from one rotation of the mirror, so the deviation made by each surface distortion can be evaluated.
Set up
If the Polygon scanner mirror is fixed with any tilt, the reflected laser beam through the V-slit is modulated by the tilt.
TA320 can analyze the tilt by measuring the pulse width.
TA320 Setting
Der Zeitintervall-Analysator TA320 erlaubt die kontinuierliche Messung von Periode, Impulsbreite und Zeitintervall eines Signals mit einer maximalen Geschwindigkeit von 14 MS/s und einer maximalen Auflösung von 100 Pikosekunden. Der große LCD-Touchscreen-Bildschirm erleichtert die intuitive, einfache Durchführung der Messungen und der Datenanalyse. Die hohe Bildschirm-Aktualisierungsrate ermöglicht eine unmittelbare Anzeige der Analyse-Ergebnisse.